Fig. 4
Isoflurane was delivered to a section of corrugated tubing (∼150 ml, gray cylinder) in series with a pair of spring-loaded bellows (as shown on the right in the schematic), ventilated with tidal volume 500 ml and rate 20 min−1: (a) characteristic trace of flow measured for two breaths; (b) corresponding tidal volume (solid) with real-time generation of actuating signal denoting anesthetic delivery window at the end of inspiration (dashed). The delivery threshold (Vth) was set to 300 ml to account for the delay in switching between sublimbs and engaging the anesthetic vaporizer; and (c) three traces of isoflurane concentration characterizing anesthetic distribution throughout the dead space compartment. Isoflurane concentration detected at the interface between dead space and alveolar compartments was negligible.

Isoflurane was delivered to a section of corrugated tubing (∼150 ml, gray cylinder) in series with a pair of spring-loaded bellows (as shown on the right in the schematic), ventilated with tidal volume 500 ml and rate 20 min−1: (a) characteristic trace of flow measured for two breaths; (b) corresponding tidal volume (solid) with real-time generation of actuating signal denoting anesthetic delivery window at the end of inspiration (dashed). The delivery threshold (Vth) was set to 300 ml to account for the delay in switching between sublimbs and engaging the anesthetic vaporizer; and (c) three traces of isoflurane concentration characterizing anesthetic distribution throughout the dead space compartment. Isoflurane concentration detected at the interface between dead space and alveolar compartments was negligible.