Abstract

Active thermography techniques are of interest for quality assurance of additive manufacturing processes. However, accurate measurements of thermophysical properties of materials are required to successfully implement active thermography. In particular, the spectral absorption coefficient of materials commonly used in additive manufacturing must be known to accurately predict the spatial distribution of thermal energy generated from absorption of power emitted by a laser or pulsed flash lamp. Accurate measurements of these optical properties are also needed to develop greater understanding of additive manufacturing processes that rely on radiative heat transfer to fuse powders. This paper presents spectral absorption coefficient measurements and uncertainty estimates of fully and partially dense ABS, PLA, and Polyamide 12 samples.

Introduction

Additive manufacturing (AM) has potential benefits in geometric freedom and reduced lead times but is subject to various sources of defects [1]. Techniques are needed to detect these defects during production. Currently, researchers seek to use active thermography to detect defects during AM processes [24].

Existing thermography models use transient surface temperature measurements to predict the existence and location of defects [58]. The optical and thermophysical properties of the material are assumed to be known in these models, so accurate values are required to successfully determine whether defects exist and to estimate their location. These models also depend on approximations regarding the depth of the region in which the incident radiative pulse is converted to thermal energy within the sample. The materials analyzed with these models are often approximated as highly absorbing (κλ → ∞), so the incident pulse is assumed to be completely absorbed within an infinitesimal layer at the surface [58].

Transient surface temperature measurements used in thermography models depend significantly on the depth at which radiation is absorbed. Thus, thermography models in which approximate radiation being absorbed entirely at the surface are not suitable for the analysis of materials that have large absorption depths. However, thermography models that account for in-depth absorption can be used to predict the transient temperature of highly absorptive, even semitransparent material [9].

The relationship between in-depth absorption and transient surface temperature is illustrated in Fig. 1 by comparing data retrieved from a numerical simulation. This simulation models the response of an ABS sample irradiated with a short pulse when the spectral absorption coefficient, κλ, was set to 0.025 μm−1 and when it approached infinity. The spectral absorption coefficient determines the exponential rate at which spectral intensity decreases as it propagates into a material. Thus, it is used to determine the depth of the region in which radiative heating occurs. The simulation results demonstrate that surface temperature profiles depend significantly on whether in-depth absorption is included in a thermography model or not.

Fig. 1
Surface temperature response of an ABS sample irradiated at 660 nm with 6000 W/m2 for 0.1 s. The data were retrieved from a 1D COMSOL simulation modeled with the properties of ABS (ρ = 1020 kg/m3, cp = 1386 J/(kg K), and k = 0.2256 W/(m K)). The 791,000 element, 8-mm-thick model was thermally insulated on all sides except for the top, which was irradiated and convectively cooled (h = 10 W/(K m2)). Transient surface temperatures are presented when κλ = 0.025 μm−1 and when κλ → ∞.
Fig. 1
Surface temperature response of an ABS sample irradiated at 660 nm with 6000 W/m2 for 0.1 s. The data were retrieved from a 1D COMSOL simulation modeled with the properties of ABS (ρ = 1020 kg/m3, cp = 1386 J/(kg K), and k = 0.2256 W/(m K)). The 791,000 element, 8-mm-thick model was thermally insulated on all sides except for the top, which was irradiated and convectively cooled (h = 10 W/(K m2)). Transient surface temperatures are presented when κλ = 0.025 μm−1 and when κλ → ∞.
Close modal

Thermography models typically neglect in-depth absorption effects and assume that radiation is absorbed entirely at the surface (κλ → ∞) [58]. Since the absorption coefficients of metals are large, surface absorption may be a reasonable assumption for metal AM processes. However, the absorption coefficients of polymers commonly used in AM processes are relatively low [10,11]. This increased absorption depth affects the surface temperature profiles used in typical thermography models and can result in inaccurate predictions regarding the existence and location of defects. Therefore, in-depth absorption effects should be included when active thermography is applied to polymer AM processes.

The objective of this paper is to present spectral absorption coefficients for some common AM polymers at wavelengths that are used in additive manufacturing for sintering material and performing quality control with active thermography. Traditionally, CO2 lasers have been used in AM to sinter material by emitting infrared radiation for short times (∼1 ms). However, researchers have recently demonstrated that sintering material with radiation in the visible spectrum over longer times (∼1 s) can improve the mechanical properties of the final part [12,13]. The metal halide bulb used to sinter polymer in these studies has also been used to perform active thermography [2,3]. Because of this, absorption coefficients will be presented over the spectrum of a metal halide bulb as well as two specific wavelengths within the visible spectrum. Spectral absorption coefficient measurements for AM polymers have already been taken in the terahertz or ultraviolet regime [10,11,14], but the measurements presented in this paper are unique because they primarily lie in the visible spectrum. It is anticipated that these measurements will also increase the understanding of the differences between laser sintering, multi-jet fusion, and projection-sintering AM processes.

Methods

A mathematical expression for the spectral absorption coefficient of an AM polymer is obtained by analyzing the scenario illustrated in Fig. 2.

Fig. 2
Illustration of the spectral intensity irradiating a black AM polymer fused to a glass slide. The indicated system (dotted rectangle) excludes intensity reflected off Interface 1 and includes intensity reflected off Interface 3.
Fig. 2
Illustration of the spectral intensity irradiating a black AM polymer fused to a glass slide. The indicated system (dotted rectangle) excludes intensity reflected off Interface 1 and includes intensity reflected off Interface 3.
Close modal
In this specific scenario, a black AM polymer (acrylonitrile butadiene styrene (ABS), polylactic acid (PLA), or Polyamide 12) was fused to a glass slide. It was assumed that this system was non-scattering and non-emitting and that multiple internal reflections were negligible. Justification for the assumptions and simplifications used in the model will be discussed in the Methods section under the subheading, “Justification of Assumptions and Simplifications.” The spectral absorption coefficient of the AM polymer shown in Fig. 2 was defined by Eq. (1) [15]. A complete derivation of this expression is provided in the Supplemental Material
κλp=ln((1ρλp)τλgτλgp)tp
(1)

Therefore, spectral absorption coefficients of AM polymers fused to glass slides may be calculated using measurements of the spectral, normal–normal transmissivity of the polymer–glass sample (τλgp), the spectral, normal–normal transmissivity of the glass slide (τλg), the spectral, normal–normal reflectivity of the polymer (ρλp), and the thickness of the polymer (tp). A Labsphere RT-060 integrating sphere and a micrometer were used to perform these measurements as described below.

Transmissivity Measurements.

The integrating sphere can be used to measure the spectral, normal-hemispherical transmissivity without any modifications. However, because the material was assumed to be non-scattering, the spectral, normal-hemispherical transmissivity measured with the integrating sphere and the spectral, normal–normal transmissivity used in the absorption coefficient calculations are equivalent. The process for measuring the spectral, normal-hemispherical transmissivity was as follows.

Radiation from a laser or a projector (∼9 mm diameter beam) was directed through the aperture of the integrating sphere. This light was reflected off the inner surface of the sphere multiple times until absorbed by a silicon photodiode. The current generated by the photodiode was then measured using a multimeter with a 0.1-nA resolution. These current readings are proportional to the power incident on the detector.

Samples of various compositions and thicknesses were then placed in front of the aperture. The ratio between the photodiode current when the radiation was obstructed by a sample and the photodiode current when the radiation entered the sphere freely gives the normal-hemispherical transmissivity of the sample [16]. This process is illustrated in Fig. 3.

Fig. 3
Process for spectral, normal-hemispherical transmissivity measurements using an integrating sphere. (a) Radiation enters the sphere unobstructed and current is measured from a photodiode. (b) Radiation is obstructed by a sample and current is measured from a photodiode.
Fig. 3
Process for spectral, normal-hemispherical transmissivity measurements using an integrating sphere. (a) Radiation enters the sphere unobstructed and current is measured from a photodiode. (b) Radiation is obstructed by a sample and current is measured from a photodiode.
Close modal
Dark current effects were subtracted from the measurements as shown in Eq. (2). Subscripts “m” and “d” indicate whether the light source was on or off while the photodiode current was measured. The numerator of this equation represents the difference between the currents measured from the photodiode when the light source was on and off, while a sample was mounted in front of the aperture as shown in Fig. 3(b). The denominator contains photodiode current measurements when the aperture was left uncovered as shown in Fig. 3(a). As previously stated, the material was assumed to be non-scattering; thus, the spectral, normal-hemispherical transmissivity and the spectral, normal–normal transmissivity are both described by Eq. (2)
formula
(2)

Reflectivity Measurements.

The integrating sphere can be used to measure the spectral, directional-hemispherical reflectivity as illustrated in Fig. 4 [16]. The spectral, normal–normal reflectivity used to calculate the spectral absorption coefficient was approximated by using these spectral, directional-hemispherical reflectivity measurements. The validity of this approximation is discussed in the Methods section under the subheading “Justification of Assumptions and Simplifications.”

Fig. 4
Process for spectral, directional-hemispherical reflectivity measurements using an integrating sphere. (a) Radiation enters the sphere unobstructed and is partially reflected off a calibration standard, while the resulting diode detector current is measured. (b) The calibration standard is replaced with a sample and the process is repeated.
Fig. 4
Process for spectral, directional-hemispherical reflectivity measurements using an integrating sphere. (a) Radiation enters the sphere unobstructed and is partially reflected off a calibration standard, while the resulting diode detector current is measured. (b) The calibration standard is replaced with a sample and the process is repeated.
Close modal

The process for measuring the spectral, directional-hemispherical reflectivity was similar to that of the spectral, normal-hemispherical transmissivity with two differences. First, the sample shown in Fig. 3(b) was moved to the opposite side of the sphere, so the collimated beam (∼9 mm diameter) entered the sphere unattenuated. Second, a reflectance calibration standard (ρλstd ≈ 0.989 +/−0.004) was used for baseline measurements. The experimental setup for the spectral, directional-hemispherical reflectivity measurements is illustrated in Fig. 4. Orienting the beam as shown in Fig. 4 ensures that specular reflections are captured within the sphere.

Like Eq. (2), dark current effects were subtracted from the reflectivity measurements in Eq. (3). The numerator in this equation is comprised of current measurements when the sample was mounted as shown in Fig. 4(b), whereas the denominator contains current measurements when the calibration standard was mounted as shown in Fig. 4(a). The reflectivity of the calibration standard is represented by ρλstd. As previously explained, the spectral, normal–normal reflectivity used to calculate the spectral absorption coefficient was approximated by using the spectral, directional-hemispherical reflectivity measurements
formula
(3)

Thickness and Density Measurements.

Sample thickness was measured using an analog micrometer with a 1.3-μm resolution. Three measurements were averaged together at randomly selected locations around the midpoint of the sample to obtain the reported thicknesses shown in Table 3 in the Results section.

Sample density was determined using Archimedes’ principle by measuring the sample weight before and after submersion in Isopropanol with a scale that has a resolution of 0.0002 g. The exact process is described thoroughly in the literature [17]. Sample densities are listed with the spectral absorption coefficients of the partially dense samples in Table 3.

Uncertainty Analysis.

Uncertainty estimations for the sample thickness, transmissivity, and reflectivity measurements were combined together to obtain the uncertainties of the spectral absorption coefficient calculations. As explained previously, three thickness measurements were averaged together for each sample. The standard deviation of these three measurements was multiplied by the 95%, two-tailed, student’s t-table value to estimate the uncertainties of the averaged sample thickness measurements, Ut.

The uncertainties of the transmissivity measurements, Uτ, were estimated using Eq. (4)
formula
(4)
where UI is the uncertainty of the current measured from a photodiode using a multimeter for the four different configurations indicated by subscripts m, d, s, and e found in Eq. (2). For each configuration, UI was found using Eq. (5)
UI=Uo2+Ui2+Unoise2
(5)
where Uo is the resolution of the multimeter and was equal to 0.1 nA. Unoise is uncertainty due to measurement noise and was estimated as follows. For each current measurement described in Eq. (2), 30 samples were taken. The standard deviation between these samples was multiplied by the 95%, two-tailed, student’s t-table value to obtain Unoise. Ui is the instrument uncertainty and was defined by Eq. (6)
Ui=±tv,95*Sxy
(6)
where Sxy is the standard error of the instrument calibration curve. A linear calibration curve was found by measuring the current output of the photodiode as a function of the input laser power. The standard error was multiplied by the 95%, two-tailed, student’s t-table value to obtain an instrument uncertainty of Ui = +/−8.857 nA.
The uncertainties of the reflectivity measurements Uρ were estimated using Eq. (7)
formula
(7)
where UI is the uncertainty of the current measured from a photodiode using a multimeter for the four different configurations indicated by subscripts m, d, s, and c found in Eq. (3). URstd is the uncertainty of the calibrated reflectance standard and is tabulated by the manufacturer as a function of the wavelength of incident irradiation and the reflectivity of the sample being measured. The range of URstd used in in this project was between 0.0022 and 0.0064.
The preceding uncertainty estimations were combined in Eq. (8) to obtain the uncertainties of the spectral absorption coefficient calculations, Uκ, for each sample individually
Uκ=(κλρλ*Uρ)2+(κλτλ*Uτ)2+(κλt*Ut)2
(8)
Equation (9) was used to estimate the uncertainty associated with averaging multiple spectral absorption coefficient calculations together
Uκ¯=2[Uκo2+(Sxm)2]12
(9)
where Sx is the standard deviation between spectral absorption coefficient calculations from different samples of the same material, and m is the number of samples whose spectral absorption coefficients are being averaged together. Uκo is an uncertainty estimation using averaged inputs as shown in Eq. (10)
Uκo=(κλ¯ρλ¯*Uρ¯)2+(κλ¯τλ¯*Uτ¯)2+(κλ¯t¯*Ut¯)2
(10)
where ρλ¯ is the spectral, normal–normal reflectivity averaged between multiple samples of the same material, τλ¯ is the spectral, normal–normal transmissivity averaged between multiple samples of the same material, t¯ is the thickness averaged between multiple samples of the same material, and U¯ is the averaged uncertainty of the reflectivity, transmissivity, or thickness measurements indicated by subscripts ρ, τ, and t. κλ¯ was defined by Eq. (11)
κλ¯=ln((1ρλ¯)τλ¯)t¯
(11)

Justification of Assumptions and Simplifications.

The system was said to be non-scattering when deriving Eqs. (1)–(3). This simplification is consistent with the models used in the literature [2,3,59]. Although some scattering will occur, the measurements and calculations presented in this paper do not consider scattering effects. Further investigation is needed to understand how additively manufactured polymer parts scatter radiation.

It was assumed that the spectral, normal–normal reflectivity used to calculate the spectral absorption coefficient could be approximated with the spectral, directional-hemispherical reflectivity measurements from the integrating sphere system. This assumption was validated by comparing measurements taken with the integrating sphere system to spectral, normal–normal reflectivity calculations using the Fresnel equation and reported indices of refraction [15,18]. As shown in Table 1, the difference between the spectral, directional-hemispherical measurements and the range of spectral, normal–normal reflectivity calculations was less than 0.01. Assumptions and approximations used in this paper regarding internal reflections and reflectivity measurements will need to be altered when analyzing highly reflective (white) AM polymers.

Table 1

Comparing spectral, directional-hemispherical reflectivity measurements from an integrating sphere system to spectral, normal–normal reflectivity calculations from Fresnel’s equation

MaterialIndex of refraction [18]ρλ⊥⊥ Calculated using Fresnel’s equation [15]ρλ Measured with the integrating sphere
PLA1.35–1.450.022–0.0340.039
ABS1.540.0450.052
Polyamide 121.52–1.530.043–0.0440.037
MaterialIndex of refraction [18]ρλ⊥⊥ Calculated using Fresnel’s equation [15]ρλ Measured with the integrating sphere
PLA1.35–1.450.022–0.0340.039
ABS1.540.0450.052
Polyamide 121.52–1.530.043–0.0440.037

Reflections at the polymer–glass interface (Interface 2 in Fig. 2) were neglected because of the similar indices of refraction. PLA, ABS, and Polyamide 12 have refractive indices of about 1.35–1.45, 1.54, and 1.52–1.53, whereas the borosilicate glass has a refractive index of about 1.5 [18,19].

Reflections at the glass–air interface (Interface 3 in Fig. 2) were also neglected. Based on the spectral absorption coefficients measured in this project, transmittance through a sample was 24% at most, and reflectance off a glass slide was 8% at most. Using these values, it can be shown that less than one percent of the spectral intensity incident on the left surface in Fig. 2 would transmit through the polymer–glass sample, reflect off the glass–air interface (Interface 3), and transmit back through the sample to the left through Interface 1. Therefore, it was assumed that the spectral, normal–normal reflectivity of the polymer–glass combination was equivalent to the spectral, normal–normal reflectivity of the polymer used in the spectral absorption coefficient calculations. This assumption would not be valid if the samples were highly transparent.

Sample Creation.

Partially and fully dense samples were created from black ABS, PLA, and Polyamide 12. The fully dense samples were created using two different methods. Some were created by melting material on 1-mm-thick glass slides in an oven, while a weight (either 10 or 20 pounds) rested on their surface. Representative samples are shown in Fig. 5. Other fully dense samples were created using a LECO PR-32 heated hydraulic press. The samples created using the hydraulic press were not fused to glass slides. However, Eq. (1) was still used to calculate the spectral absorption coefficient of these samples by increasing the value of the transmissivity of the glass slide in the equation to a value of one (τλg=1).

Fig. 5
Partially dense ABS (top) and fully dense PLA and Polyamide 12 (bottom) samples
Fig. 5
Partially dense ABS (top) and fully dense PLA and Polyamide 12 (bottom) samples
Close modal

To verify that the different manufacturing methods did not impact the spectral absorption coefficient calculations, several fully dense PLA samples were created using either the heated hydraulic press or the oven and weight. The calculated spectral absorption coefficients of these two different sets of samples differed by less than 0.0008 μm−1. Therefore, any variations due to differences in the manufacturing process were considered negligible.

The partially dense PLA and ABS samples were created via fused filament fabrication (FFF) with a CreatorPro FlashForge 3D printer set at full infill density. The black PLA filament was purchased from Hatchbox, while the black ABS filament was purchased from IC3D. These 50 by 25 mm samples had various thicknesses based on the number of layers deposited during manufacturing as shown in Table 3. The partially dense Polyamide 12 samples were created using a large area projection-sintering (LAPS) system [12,13]. These 15 by 20 mm samples were created from PA2202 black Polyamide 12 powder purchased from EOS. None of the partially dense samples were fused to glass slides; therefore, the transmissivity of the glass slide used in Eq. (1) was set equal to one when calculating their spectral absorption coefficients.

Projector Spectrum Measurement.

As mentioned in the Introduction, the literature utilizes existing equipment (including a modified projector with a metal halide bulb) for both sintering material and performing active thermography. For this purpose, the spectrum of the X316 Optoma Projector was measured using a spectrometer [20]. The projector was previously modified for use in large area projection sintering (LAPS) [12,13]. Figure 6 shows the measured spectrum, which has been normalized. The uncertainty envelope was determined by taking 100 samples at each wavelength over the spectrum and multiplying the standard deviation between these measurements by the 95%, two-tailed, student’s t-table value. The maximum of these calculated values was designated as the uncertainty of the system.

Fig. 6
Measured spectrum of the X316 Optoma Projector
Fig. 6
Measured spectrum of the X316 Optoma Projector
Close modal

Results

Measurements and Calculations.

Spectral absorption coefficients of PLA, ABS, and Polyamide 12 at 532 nm, 660 nm, and over the entire spectrum shown in Fig. 6 are presented in Tables 2 and 3.

Table 2

Absorption coefficient of fully dense AM polymers (μm−1)

MaterialColorManufacturer660 nm532 nmProjector spectrum
PLABlackHatchbox0.022 +/− 0.0020.026 +/− 0.0030.023 +/− 0.01
ABSBlackIC3D0.025 +/− 0.0030.028 +/− 0.0050.025 +/− 0.02
Polyamide 12BlackEOS0.026 +/− 0.0030.030 +/− 0.0050.023 +/− 0.02
MaterialColorManufacturer660 nm532 nmProjector spectrum
PLABlackHatchbox0.022 +/− 0.0020.026 +/− 0.0030.023 +/− 0.01
ABSBlackIC3D0.025 +/− 0.0030.028 +/− 0.0050.025 +/− 0.02
Polyamide 12BlackEOS0.026 +/− 0.0030.030 +/− 0.0050.023 +/− 0.02
Table 3

Absorption coefficient of partially dense AM polymers (μm−1)

ABS
Number of layers printed234
Thickness (μm)206 +/− 3297 +/− 6418 +/− 3
Density (kg/m3)108310771080
κλ at 532 nm (μm−1)0.021 +/− 0.0010.027 +/− 0.0040.022 +/− 0.01
κλ at 660 nm (μm−1)0.018 +/− 0.0010.023 +/− 0.0040.020 +/− 0.01
κλ over projector spectrum (μm−1)0.021 +/− 0.030.014 +/− 0.07
PLA
Number of layers printed234
Thickness (μm)189 +/− 3268 +/− 7361 +/− 3
Density (kg/m3)128912891291
κλ at 532 nm (μm−1)0.026 +/− 0.0030.028 +/ 0.0030.025 +/− 0.01
κλ at 660 nm (μm−1)0.021 +/− 0.0010.022 +/− 0.0020.020 +/− 0.01
κλ over projector spectrum m−1)0.023 +/− 0.03
Polyamide 12
Number of layers printed223
Thickness (μm)154 +/− 12155 +/− 6239 +/− 4
Density (kg/m3)106510541065
κλ at 532 nm (μm−1)0.025 +/− 0.0030.024 +/− 0.0020.024 +/− 0.007
κλ at 660 nm (μm−1)0.022 +/− 0.0020.020 +/− 0.0010.022 +/− 0.003
κλ over projector spectrum (μm−1)0.019 +/− 0.0090.017 +/− 0.0070.016 +/− 0.016
ABS
Number of layers printed234
Thickness (μm)206 +/− 3297 +/− 6418 +/− 3
Density (kg/m3)108310771080
κλ at 532 nm (μm−1)0.021 +/− 0.0010.027 +/− 0.0040.022 +/− 0.01
κλ at 660 nm (μm−1)0.018 +/− 0.0010.023 +/− 0.0040.020 +/− 0.01
κλ over projector spectrum (μm−1)0.021 +/− 0.030.014 +/− 0.07
PLA
Number of layers printed234
Thickness (μm)189 +/− 3268 +/− 7361 +/− 3
Density (kg/m3)128912891291
κλ at 532 nm (μm−1)0.026 +/− 0.0030.028 +/ 0.0030.025 +/− 0.01
κλ at 660 nm (μm−1)0.021 +/− 0.0010.022 +/− 0.0020.020 +/− 0.01
κλ over projector spectrum m−1)0.023 +/− 0.03
Polyamide 12
Number of layers printed223
Thickness (μm)154 +/− 12155 +/− 6239 +/− 4
Density (kg/m3)106510541065
κλ at 532 nm (μm−1)0.025 +/− 0.0030.024 +/− 0.0020.024 +/− 0.007
κλ at 660 nm (μm−1)0.022 +/− 0.0020.020 +/− 0.0010.022 +/− 0.003
κλ over projector spectrum (μm−1)0.019 +/− 0.0090.017 +/− 0.0070.016 +/− 0.016

Note: The black ABS, PLA, and Polyamide 12 materials were purchased from IC3D, Hatchbox, and EOS, respectively. FFF was used to create the ABS and PLA samples, whereas LAPS was used to create the Polyamide 12 samples.

Validation.

The reflectivity measurements were validated as shown in Table 1, using Fresnel’s equation [15]. The transmissivity measurement process was validated by comparing measurements from the integrating sphere system to measurements from a Cary-60 spectrophotometer. This spectrophotometer was designed to measure the absorptance of liquid solutions but was modified to hold and measure the transmissivity of one PLA sample. The integrating sphere system was also modified to replicate the geometric configuration of the spectrophotometer. Figure 7 illustrates how these two measurement systems validate each other.

Fig. 7
Comparing spectral, normal–normal transmissivity measurements obtained using a spectrophotometer and an integrating sphere for a black, fully dense, 109-μm-thick, PLA sample fused to a glass slide
Fig. 7
Comparing spectral, normal–normal transmissivity measurements obtained using a spectrophotometer and an integrating sphere for a black, fully dense, 109-μm-thick, PLA sample fused to a glass slide
Close modal

Figure 8 illustrates the accuracy of the calculated absorption coefficients in predicting transmission. Exponential curves based on Beer’s Law [15] and the calculated absorption coefficients were plotted next to the spectral, normal-hemispherical transmissivities measured with the integrating sphere system. The exponential curves fit the transmission data well.

Fig. 8
Transmissivity of (a) black, fully dense PLA, (b) ABS and (c) Polyamide 12 samples fused to glass slides, irradiated at 660 nm
Fig. 8
Transmissivity of (a) black, fully dense PLA, (b) ABS and (c) Polyamide 12 samples fused to glass slides, irradiated at 660 nm
Close modal

Conclusions

Spectral absorption coefficient calculations and uncertainty estimations have been presented for black PLA, ABS, and Polyamide 12 samples. The reflectivity measurements used in the absorption coefficient calculations were validated using Fresnel’s equation. The transmissivity measurements used in the absorption coefficient calculations have been validated by comparing results obtained from a system based on an integrating sphere, and a spectrophotometer.

While the data reveal that PLA is most sensitive to effects from in-depth absorption, all three polymers have measured spectral absorption coefficients in the visible spectrum between 0.02 and 0.03 μm−1. At a depth of 96 μm, approximately 90% of irradiation in the visible spectrum incident on the surface of a PLA part is absorbed. If a defect is present at this depth, it is likely that a significant portion of the remaining irradiation will be transmitted through the defect and into the material below. In parts made with ABS and Polyamide 12, the depth at which 90% of the irradiation is absorbed decreases from 96 to 87 and 82 μm, respectively. Transmission through subsurface defects will distort surface temperatures necessary for active thermography measurements. Thermographic models should include effects from in-depth absorption when analyzing defects in AM polymers less than 100 μm below the surface of the part.

These results also have important implications for sintering materials with radiation in the visible spectrum. The absorption depths mentioned in the previous paragraph are significant relative to typical layer thickness of many AM processes (∼50–100 μm). Traditionally, powder-bed fusion AM processes have used infrared radiation from CO2 lasers to sinter material. However, using radiation in the visible spectrum would result in different absorption profiles which may contribute to differences in the properties of the resulting parts [12].

Future work is needed to extend these measurements to different AM materials, colored filaments, and spectral regions. Assumptions and approximations used in this paper regarding internal reflections and reflectivity measurements will need to be altered when analyzing white AM polymers. Understanding the spectral absorption of AM materials will give researchers and developers the necessary tools to improve AM processes that utilize radiative heat transfer to sinter powders. In addition, accurate spectral absorption coefficient measurements are necessary to successfully adapt active thermography techniques as a non-destructive, non-contact means of in situ monitoring of AM processes.

Acknowledgment

We would like to thank Trevor Boardman for providing the partially dense ABS and PLA samples, as well as Spencer Abbot for creating the partially dense Polyamide 12 samples. In addition, we would like to acknowledge Dr. William G. Pitt for providing the spectrophotometer and Dr. Troy Munro for providing the lasers and spectrometer to perform these measurements. Finally, we would like to acknowledge Chandler Warr and Derek Sanchez for their assistance in experimental setup.

Data Availability Statement

The data sets generated and supporting the findings of this article are obtainable from the corresponding author upon reasonable request.

Funding Data

  • This research was funded in part by the Utah NASA Space Grant Consortium, NASA Grant #NNX15AI24H, and by the National Science Foundation through award CMMI- 1851728.

Nomenclature

h =

coefficient of convection

k =

thermal conductivity

m =

number of measurements averaged together

s =

distance into the sample from its surface in the direction of the irradiation

t =

sample thickness; (tp + tg)

U =

measurement uncertainty

t¯ =

thickness measurements averaged between multiple samples

U¯ =

measurement uncertainty averaged between multiple samples

tg =

thickness of the glass slide

tp =

thickness of the polymer sample

tv,95 =

95%, two-tailed, student’s t-table value

Iλ =

spectral intensity

Iλo =

incident spectral intensity (Iλ(0))

Id =

current measured from the photodiode when the radiation source was off

Im =

current measured from the photodiode when the radiation source was on

Sx =

standard deviation

Sxy =

standard error of the instrument calibration curve

Ui =

instrument uncertainty

UI =

uncertainty of the current measured from the photodiode

Uko =

uncertainty estimation using averaged inputs

Unoise =

uncertainty due to measurement noise

Uo =

resolution of the multimeter used to measure current

cp =

specific heat capacity

κλ =

spectral absorption coefficient

κλg =

spectral absorption coefficient of the glass slide

κλp =

spectral absorption coefficient of the polymer sample

κλ¯ =

spectral absorption coefficient averaged between multiple samples

ρ =

density

ρλstd =

spectral reflectivity of the calibration standard

ρλ =

spectral, normal–normal reflectivity

=

spectral, directional-hemispherical reflectivity

ρλp =

spectral, normal–normal reflectivity of the polymer sample

ρλgp =

spectral, normal–normal reflectivity of the polymer–glass combination

ρλ¯ =

spectral, normal–normal reflectivity averaged between multiple samples

τλ =

spectral, normal–normal transmissivity

=

spectral, normal-hemispherical transmissivity

τλg =

spectral, normal–normal transmissivity of glass slide

τλp =

spectral, normal–normal transmissivity of polymer sample

τλgp =

spectral, normal–normal transmissivity of polymer–glass combination

τλ¯ =

spectral, normal–normal transmissivity averaged between multiple samples

Subscripts

c =

current measured from the photodiode when radiation entered the integrating sphere and a calibration standard was mounted as shown in Fig. 4(a) 

e =

current measured from the photodiode when radiation entered the integrating sphere unobstructed as shown in Fig. 3(a) 

s =

current measured from the photodiode when radiation entered the integrating sphere and a sample was mounted as shown in either Fig. 3(b) or Fig. 4(b) 

References

1.
Yao
,
B.
,
Imani
,
F.
,
Sakpal
,
A. S.
,
Reutzel
,
E. W.
, and
Yang
,
H.
,
2018
, “
Multifractal Analysis of Image Profiles for the Characterization and Detection of Defects in Additive Manufacturing
,”
ASME J. Manuf. Sci. E.
,
140
(
3
), p.
031014
. 10.1115/1.4037891
2.
Pierce
,
J.
, and
Crane
,
N. B.
,
2019
, “
Impact of Pulse Length on the Accuracy of Defect Depth Measurements in Pulse Thermography
,”
ASME J. Heat Transfer
,
141
(
4
), p.
042002
. 10.1115/1.4042785
3.
Pierce
,
J. R.
, and
Crane
,
N. B.
,
2017
, “
Preliminary Nondestructive Testing Analysis on 3D Printed Structure Using Pulsed Thermography
,”
Proceedings of the ASME 2017 International Mechanical Engineering Congress and Exposition. Volume 8: Heat Transfer and Thermal Engineering.
,
Tampa, FL
,
Nov. 3–9
, ASME, p. V008T10A083. https://doi.org/10.1115/IMECE2017-71935
4.
Metz
,
C.
,
Franz
,
P.
,
Fischer
,
C.
,
Wachtendorf
,
V.
, and
Maierhofer
,
C.
,
2019
, “
Active Thermography for Quality Assurance of 3D-Printed Polymer Structures
,”
Quant. Infrared Thermogr. J.
10.1080/17686733.2019.1686896
5.
Almond
,
D.
, and
Pickering
,
S.
,
2012
, “
An Analytical Study of the Pulsed Thermography Defect Detection Limit
,”
J. Appl. Phys.
,
111
(
9
), p.
093510
. 10.1063/1.4704684
6.
Ringermacher
,
H.
,
Archacki
,
R.
, and
Veronesi
,
W.
,
1998
, “
Nondestructive Testing: Transient Depth Thermography
,”
U.S. Patent No. 5,711,603
.
7.
Maldague
,
X.
, and
Marinetti
,
S.
,
1996
, “
Pulse Phase Infrared Thermography
,”
J. Appl. Phys.
,
79
(
5
), pp.
2694
2698
. 10.1063/1.362662
8.
Busse
,
G.
,
Wu
,
D.
, and
Karpen
,
W.
,
1992
, “
Thermal Wave Imaging With Phase Sensitive Modulated Thermography
,”
J. Appl. Phys.
,
71
(
8
), pp.
3962
3965
. 10.1063/1.351366
9.
Bernegger
,
R.
,
Altenburg
,
S. J.
, and
Maierhofer
,
C.
,
2020
, “
Quantification of Delaminations in Semitransparent Solids Using Pulsed Thermography and Mathematical 1D Models
,”
Int. J. Thermophys.
,
41
(
67
).10.1007/s10765-020-02642-7.
10.
Barzic
,
A. I.
, and
Barzic
,
R. F.
,
2015
, “
Optical Properties of Polysaccharide/Polylactide Blends
,”
UPB Sci. Bull.
,
77
(
4
), pp.
293
302
.
11.
Squire
,
A. D.
, and
Lewis
,
R. A.
,
2017
, “
Mechanical and Optical Viability of Eighteen Filaments for 3D Printing of Terahertz Components
,”
Proceedings of the 42nd International Conference on Infrared, Millimeter, and Terahertz Waves
,
Cancun, MX
,
Aug. 27, 2017
, pp.
1
2.
12.
Craft
,
G.
,
Nussbaum
,
J.
,
Crane
,
N.
, and
Harmon
,
J. P.
,
2018
, “
Impact of Extended Sintering Times on Mechanical Properties in PA-12 Parts Produced by Powderbed Fusion Processes
,”
Addit. Manuf.
,
22
, pp.
800
806
. 10.1016/j.addma.2018.06.028
13.
Nussbaum
,
J.
, and
Crane
,
N.
,
2018
, “
Evaluation of Processing Variables in Polymer Projection Sintering
,”
Rapid Prototyping J.
,
24
(
5
), pp.
880
885
. 10.1108/RPJ-04-2017-0070
14.
Elvidge
,
C. D.
,
Keith
,
D. M.
,
Tuttle
,
B. T.
, and
Baugh
,
K. E.
,
2010
, “
Spectral Identification of Lighting Type and Character
,”
Sensors
,
10
(
4
), pp.
3961
3988
. 10.3390/s100403961
15.
Modest
,
M.
,
2013
,
Radiative Heat Transfer
,
Academic Press
,
NY
,
Chap. 3,9
.
16.
Integrating Sphere Theory and Applications
,
Labsphere
,
North Sutton, NH
, https://www.labsphere.com/site/assets/files/2551/integrating_sphere_theory_apps_tech_guide.pdf, Accessed August 11, 2020.
17.
Density Determination Kit Instruction Manual
,
OHAUS Cooperation
,
Parsippany, NJ
, https://dmx.ohaus.com/WorkArea/showcontent.aspx?id=25404, Accessed July 28, 2020.
18.
Wypych
,
G.
,
2016
,
Handbook of Polymers
,
ChemTec Publishing
,
Toronto
.
19.
Mertens
,
H.
, and
Polman
,
A.
,
2006
, “
Depth-Resolved Nanostructure and Refractive Index of Borosilicate Glass Doped With Ag Nanocrystals
,”
Opt. Mater.
,
29
(
2–3
), pp.
326
331
. 10.1016/j.optmat.2005.11.002
20.
GREEN-Wave Miniature Fiber Optic Spectrometers for VIS-NIR & OEM
,
StellarNet Inc
,
Tampa, FL
, https://www.stellarnet.us/wp-content/uploads/StellarNet-GREEN-Wave-SPEC.pdf, Accessed July 28, 2020.

Supplementary data