Valvular heart disease is a significant problem. The primary case physician initially does assessment through auscultation. Accuracy in classification of sounds is suboptimal (20 to 40%). Lower frequencies of heart sounds are important in classification of murmurs associated with valvular heart disease. We find stethoscope sound intensity capture falls significantly at the 1500 Hz range and lower. Strategies to improve auscultation accuracy include improving stethoscope features or developing a device that, when used with the stethoscope, augments sound capturing abilities at lower frequencies. Testing necessitated development of a reliable (without significant intra-sound variation) cardiac sound reproduction device. A sound permeable contracting polymer when used with a stethoscope signficantly increases sound intensity captured in the 625 Hz to 1500 Hz range, when tested with a reliable cardiac sound reproduction device. We prepared an air-sealed device with an amplifier, four internal speakers capable of emitting high quality, low frequency sounds, and a listening pad. An existing electronic stethoscope with, and then without, a sound permeable contracting polymer captured three sounds (normal, innocent systolic murmur, pathological systolic murmur) five times per sound. The sounds were placed in computer files. FFTs were constructed. Sound intensity within the 625 Hz to 1400 Hz range, when the sound permeable contracting polymer is used with the electronic stethoscope, relatively improves, on average, approximately 11 dB, compared to sound captured with the same electronic stethoscope without the sound permeable contracting polymer. This difference is numerically statistically significant . Intra-sound variability testing (standard deviation) of FFTs was not significant. A sound permeable contracting polymer used with an electronic recording stethoscope significantly improves sound intensity in an important auscultation frequency range. Intra-sound testing variation was insignificant. A study is underway to demonstrate impact using an absolute reference point. However, as amplification within existing electronic stethoscopes is commercially available, potential variation in relative reference points may be overcome with existing amplification features on electronic stethoscopes, allowing improved capture of heart sounds within the 625 Hz to 1400 Hz range. Limitations include the need for human subject study. Further testing, including human subjects testing is needed prior to application for FDA approval. FDA approval is needed before any use on humans. The methodology should not be utilized clinically without FDA approval.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
A Cardiac Sound Reproduction Apparatus for Improved Stethoscope Testing
G. Nordehn,
G. Nordehn
University of Minnesota Medical School Duluth
Search for other works by this author on:
R. Ulseth,
R. Ulseth
Itasca Community College
, Engineering Department
Search for other works by this author on:
T. Bartek,
T. Bartek
Itasca Community College
, Engineering Department
Search for other works by this author on:
B. McDonald,
B. McDonald
Itasca Community College
, Engineering Department
Search for other works by this author on:
A. McNally,
A. McNally
Itasca Community College
, Engineering Department
Search for other works by this author on:
M. Menart,
M. Menart
Itasca Community College
, Engineering Department
Search for other works by this author on:
S. Eames,
S. Eames
Itasca Community College
, Engineering Department
Search for other works by this author on:
M. Flynn,
M. Flynn
Itasca Community College
, Engineering Department
Search for other works by this author on:
B. Blaire,
B. Blaire
Itasca Community College
, Engineering Department
Search for other works by this author on:
A. Drewek,
A. Drewek
Itasca Community College
, Engineering Department
Search for other works by this author on:
E. Branson,
E. Branson
University of Minnesota Duluth
, Department of Electrical and Computer Engineering
Search for other works by this author on:
S. Burns
S. Burns
University of Minnesota Duluth
, Department of Electrical and Computer Engineering
Search for other works by this author on:
G. Nordehn
University of Minnesota Medical School Duluth
R. Ulseth
Itasca Community College
, Engineering Department
T. Bartek
Itasca Community College
, Engineering Department
B. McDonald
Itasca Community College
, Engineering Department
A. McNally
Itasca Community College
, Engineering Department
M. Menart
Itasca Community College
, Engineering Department
S. Eames
Itasca Community College
, Engineering Department
M. Flynn
Itasca Community College
, Engineering Department
B. Blaire
Itasca Community College
, Engineering Department
A. Drewek
Itasca Community College
, Engineering Department
E. Branson
University of Minnesota Duluth
, Department of Electrical and Computer Engineering
S. Burns
University of Minnesota Duluth
, Department of Electrical and Computer EngineeringJ. Med. Devices. Jun 2009, 3(2): 027549 (1 pages)
Published Online: July 24, 2009
Article history
Published:
July 24, 2009
Citation
Nordehn, G., Ulseth, R., Bartek, T., McDonald, B., McNally, A., Menart, M., Eames, S., Flynn, M., Blaire, B., Drewek, A., Branson, E., and Burns, S. (July 24, 2009). "A Cardiac Sound Reproduction Apparatus for Improved Stethoscope Testing." ASME. J. Med. Devices. June 2009; 3(2): 027549. https://doi.org/10.1115/1.3147377
Download citation file:
354
Views
Get Email Alerts
Cited By
Design, Development and Validation of a Smart Cochlear Three-Dimensional-Printed Model to Train ENT Surgeons
J. Med. Devices (December 2023)
Experimental Investigation of the Calcified Plaque Material Removal Rate in Coronary Rotational Atherectomy
J. Med. Devices (December 2023)
A Novel Design Method for the Knee Joint of the Exoskeleton Based on the Modular Wearable Sensor
J. Med. Devices (December 2023)
Related Articles
Integration of a Quantitative Feedback Theory (QFT)-Based Active Noise Canceller and 3D Audio Processor to Headsets
J. Vib. Acoust (October,2007)
The Evolution of the External Left Ventricular Assist Device
J. Med. Devices (June,2010)
Design Optimization of Single-Port Minimally Invasive Intervention Devices
J. Med. Devices (June,2009)
Related Chapters
Synthesis and Characterization of Carboxymethyl Chitosan Based Hybrid Biopolymer Scaffold
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3
The Effect of Temperature on Polymer Supported Titanium Dioxide Photocatalyst for Degradation of Volatile Organic Compounds
International Conference on Computer and Electrical Engineering 4th (ICCEE 2011)
LC-MS/MS on Microfluidic Device for Chloramphenicol Determination in Milk and Honey Samples Based on Molecular Imprinted Polymers
International Conference on Computer and Electrical Engineering 4th (ICCEE 2011)