Abstract
Alternatives to the current invasive rib implants are discussed as from the current solutions with their constraints and disadvantages mainly in surgery application and clinical consequences. A novel rib fixator is presented with an experimental validation and characterization for a potential implementation in facilitating osteosynthesis of multifractured ribs. Testing is designed with lab facilities replicating normal breathing with the possibility of simulating cough too. Results of lab test with a ribolution rib fixator prototype are discussed from experiences with pig ribs with satisfactory results in behavior and numerical values.
Issue Section:
Research Papers
References
1.
Atta
,
H. M.
, 1999
, “
Edwin Smith Surgical Papyrus: The Oldest Known Surgical Treatise
,” Am. Surg.
,
65
(12
), p. 10597074
.https://www.researchgate.net/publication/12704490_Edwin_Smith_Surgical_Papyrus_The_oldest_known_surgical_treatise2.
Standring
,
S.
, ed., 2014
, Gray's Anatomy: The Anatomy Principles for Clinic Practice, Edra Feltrinelli
,
Milan
(Italian Edition).3.
Hall
,
J. E.
, and
Hall
,
M. E.
, 2020
, Guyton and Hall Textbook of Medical Physiology
,
Elsevier
, Amsterdam, The Netherlands.4.
Lafferty
,
P. M.
,
Anavian
,
J.
,
Will
,
R. E.
, and
Cole
,
P. A.
, 2011
, “
Operative Treatment of Chest Wall Injuries: Indications, Technique, and Outcomes
,” J. Bone Jt. Surg.
,
93
(1
), pp. 97
–110
.10.2106/JBJS.I.006965.
Chrysou
,
K.
,
Halat
,
G.
,
Hoksch
,
B.
,
Schmid
,
R. A.
, and
Kocher
,
G. J.
, 2017
, “
Lessons From a Large Trauma Center: Impact of Blunt Chest Trauma in Polytrauma Patients—Still a Relevant Problem?
,” Scand. J. Trauma, Resusc. Emerg. Med.
,
25
(1
), p. 42
.10.1186/s13049-017-0384-y6.
Connolly
,
L. P.
, and
Connolly
,
S. A.
, 2004
, “
Rib Stress Fractures
,” Clinical Nucl. Med.
,
29
(10
), pp. 614
–616
.10.1097/00003072-200410000-000027.
Sirmali
,
M.
,
Türüt
,
H.
,
Topçu
,
S.
,
Gülhan
,
E.
,
Yazici
,
U.
,
Kaya
,
S.
, and
Taştepe
,
I.
, 2003
, “
A Comprehensive Analysis of Traumatic Rib Fractures: Morbidity, Mortality and Management
,” Eur. J. Cardio-Thorac. Surg.
,
24
(1
) pp. 133
–138
.10.1016/S1010-7940(03)00256-28.
Zhang
,
Q.
,
Song
,
L.
,
Ning
,
S.
,
Xie
,
H.
,
Li
,
N.
, and
Wang
,
Y.
, 2019
, “
Recent Advances in Rib Fracture Fixation
,” J. Thorac. Disease
,
11
(S8
), pp. S1070
–S1077
.10.21037/jtd.2019.04.999.
Engel
,
C.
,
Krieg
,
J. C.
,
Madey
,
S. M.
,
Long
,
W. B.
, and
Bottlang
,
M.
, 2005
, “
Operative Chest Wall Fixation With Osteosynthesis Plates
,” J. Trauma: Inj., Infect., Crit. Care
,
58
(1
), pp. 181
–186
.10.1097/01.TA.0000063612.25756.6010.
Bottlang
,
M.
,
Helzel
,
I.
,
Long
,
W.
,
Fitzpatrick
,
D.
, and
Madey
,
S.
, 2010
, “
Less-Invasive Stabilization of Rib Fractures by Intramedullary Fixation: A Biomechanical Evaluation
,” J. Trauma
,
68
(5
), pp. 1218
–1224
.10.1097/TA.0b013e3181bb9df111.
Helzel
,
I.
,
Long
,
W.
,
Fitzpatrick
,
D.
,
Madey
,
S.
, and
Bottlang
,
M.
, 2009
, “
Evaluation of Intramedullary Rib Splints for Less- Invasive Stabilisation of Rib Fractures
,” Injury
,
40
(10
), pp. 1104
–1110
.10.1016/j.injury.2009.06.00412.
Vyhnánek
,
F.
,
Jírava
,
D.
,
Očadlík
,
M.
,
Škrabalová
,
D.
,
Šáber
,
M.
, and
Michal
,
P.
, 2018
, “
Innovated Judet Ribs Plates - Preclinical Study, First Clinical Experience
,” Acta Chir. Orthop. Traumatol. Cech.
,
85
(3
), pp. 226
–230
.https://pubmed.ncbi.nlm.nih.gov/30257784/13.
Gordy
,
S.
,
Fabricant
,
L.
,
Ham
,
B.
,
Mullins
,
R.
, and
Mayberry
,
J.
, 2014
, “
The Contribution of Rib Fractures to Chronic Pain and Disability
,” Am. J. Surg.
,
207
(5
), pp. 659
–663
.10.1016/j.amjsurg.2013.12.01214.
Dogrul
,
B. N.
,
Kiliccalan
,
I.
,
Asci
,
E. S.
, and
Peker
,
S. C.
, 2020
, “
Blunt Trauma Related Chest Wall and Pulmonary Injuries: An Overview
,” Chin. J. Traumatol.
,
23
(3
), pp. 125
–138
.10.1016/j.cjtee.2020.04.00315.
Ambrogi
,
V.
, and
Ceccarelli
,
M.
, 2020
, “
Fixing Plate for Osteosynthesis of Fractured Ribs
,” Patent Pending, No. 102019000005638, Italy.16.
Sommariva
,
L.
,
Arreguin
,
J. L.
,
Morales
,
C.
,
Ceccarelli
,
M.
,
Ambrogi
,
V.
, and
Puglisi
,
L.
, 2021
, “Design and Lab Experiences for a Fixator of Rib Fractures,” New Trends in Medical and Service Robotics - MESROB
,
Springer Nature Switzerland AG, Cham
, pp. 1
–9
.10.1007/978-3-030-58104-6_1817.
Mombelli
,
A.
,
Hashim
,
D.
, and
Cionca
,
N.
, 2018
, “
What is the Impact of Titanium Particles and Biocorrosion on Implant Survival and Complications? A Critical Review
,” Clin. Oral Implants Res.
,
29
(18
), pp. 37
–53
.10.1111/clr.1330518.
Majak
,
P.
, and
Næss
,
P. A.
, 2016
, “
Rib Fractures in Trauma Patients: does Operative Fixation Improve Outcome?
,” Curr. Opin. Crit. Care
,
22
(6
), pp. 572
–577
.10.1097/MCC.000000000000036419.
Pieracci
,
F. M.
, 2019
, “
Completely Thoracoscopic Surgical Stabilization of Rib Fractures: Can It Be Done and is It Worth It?
,” J. Thorac. Disease
,
11
(S8
), pp. S1061
–S1069
.10.21037/jtd.2019.01.7020.
Akil
,
A.
,
Ziegeler
,
S.
,
Reichelt
,
J.
,
Semik
,
M.
,
Müller
,
M. C.
, and
Fischer
,
S.
, 2019
, “
Rib Osteosynthesis is a Safe and Effective Treatment and Leads to a Significant Reduction of Trauma Associated Pain
,” Eur. J. Trauma Emerg. Surg.
,
45
(4
), pp. 623
–630
.10.1007/s00068-018-01062-521.
Brasel
,
K. J.
,
Guse
,
C. E.
,
Layde
,
P.
, and
Weigelt
,
J. A.
, 2006
, “
Rib Fractures: Relationship With Pneumonia and Mortality
,” Crit. Care Med.
,
34
(6
), pp. 1642
–1646
.10.1097/01.CCM.0000217926.40975.4B22.
Baiu
,
I.
, and
Spain
,
D.
, 2019
, “
Rib Fractures
,” J. Am. Med. Assoc.
,
321
(18
), p. 1836
.10.1001/jama.2019.231323.
Kiel
,
J.
, and
Kaiser
,
K.
, 2020
, “
Stress Reaction and Fractures
,” StatPearls [Internet],
StatPearls Publishing
,
Treasure Island, FL
, accessed Mar. 15, 2022, https://www.ncbi.nlm.nih.gov/books/NBK507835/24.
Thiruvenkatarajan
,
V.
,
Cruz Eng
,
H.
, and
Adhikary
,
S. D.
, 2018
, “
An Update on Regional Analgesia for Rib Fractures
,” Curr. Opin. Anesthesiol.
,
31
(5
), pp. 601
–607
.10.1097/ACO.000000000000063725.
Marco
,
C. A.
,
Sorensen
,
D.
,
Hardman
,
C.
,
Bowers
,
B.
,
Holmes
,
J.
, and
McCarthy
,
M. C.
, 2020
, “
Risk Factors for Pneumonia Following Rib Fractures
,” Am. J. Emerg. Med.
,
38
(3
), pp. 610
–612
.10.1016/j.ajem.2019.10.02126.
Sano
,
A.
,
Tashiro
,
K.
, and
Fukuda
,
T.
, 2015
, “
Cough-Induced Rib Fractures
,” Asian Cardiovasc. Thorac. Ann.
,
23
(8
), pp. 958
–960
.10.1177/021849231559689327.
Ramirez
,
O.
,
Ceccarelli
,
M.
,
Russo
,
M.
,
Christopher
,
R.
,
Torres
,
S. M.
, and
Urriolagoitia-Calderon
,
G.
, 2019
, “Experimental Dynamic Tests of Rib Implants, Advances in Italian Mechanism Science,” Proceedings of the Second International Conference of IFToMM Italy
,
Springer
,
Cham
, pp. 353
–361
.10.1007/978-3-030-03320-0_3828.
Aguilar
,
L. A.
,
Ceccarelli
,
M.
,
Torres-San-Miguel
,
C.
,
Urriolagoitia-Sosa
,
G.
, and
Urriolagoitia-Calderón
,
G.
, 2018
, Experimental Evaluation of Artificial Human Ribs, Multibody Mechatronic Systems
,
Springer, Cham
, pp. 434
–433
.29.
Farley
,
P.
,
Griffin
,
R. L.
,
Jansen
,
J. O.
, and
Bosarge
,
P. L.
, 2020
, “
Quantifying Pain Associated With Rib Fractures
,” J. Surgical Res.
,
246
, pp. 476
–481
.10.1016/j.jss.2019.09.032Copyright © 2023 by ASME