This work introduces a method for the in vivo measurement and analysis of turbulence within the leakage of a mechanical heart valve. Several analysis techniques were applied to ultrasound measurements acquired within the atrium of a pig, and error associated with these techniques was analyzed. The technique chosen applies cyclic averaging to mean and maximum velocity measurements within small, normalized phase windows to calculate Reynolds normal stresses in the direction of the ultrasound beam. Maximum shear stresses are estimated from these normal stresses using an analytical technique. The stresses observed were smaller than those reported from previous in vitro simulations.
Issue Section:
Technical Papers
1.
Hasenkam
, J. M.
et al., 1996
, “Hemodynamic evaluation of a new bileaflet valve prosthesis: an acute animal experimental study
,” J. Heart Valve Dis.
, 5
(6
), pp. 574
–580
.2.
Yoganathan, A. P., personal communication.
3.
Gross
, J. M.
et al., 1996
, “A microstructural flow analysis within a bileaflet mechanical heart valve hinge
,” J. Heart Valve Dis.
, 5
(6
), pp. 581
–590
.4.
Ellis
, J. T.
et al., 1996
, “Velocity measurements and flow patterns within the hinge region of a Medtronic Parallel bileaflet mechanical valve with clear housing
,” J. Heart Valve Dis.
, 5
(6
), pp. 591
–599
.5.
Woo
, Y. R.
, and Yoganathan
, A. P.
, 1986
, “In vitro pulsatile flow velocity and shear stress measurements in the vicinity of mechanical mitral heart valve prostheses
,” J. Biomech.
, 19
(1
), pp. 39
–51
.6.
Baldwin
, J. T.
et al., 1991
, “Mean velocities and Reynolds stresses within regurgitant jets produced by tilting disc valves
,” ASAIO Trans.
, 37
(3
), pp. M348–M349
M348–M349
.7.
Ellis
, J. T.
et al., 1996
, “An in vitro investigation of the retrograde flow fields of two bileaflet mechanical heart valves
,” J. Heart Valve Dis.
, 5
(6
), pp. 600
–606
.8.
Maymir
, J. C.
et al., 1997
, “Effects of tilting disk heart valve gap width on regurgitant flow through an artificial heart mitral valve
,” Artif. Organs
, 21
(9
), pp. 1014
–1025
.9.
Steegers
, A.
et al., 1999
, “Leakage flow at mechanical heart valve prostheses: improved washout or increased blood damage?
,” J. Heart Valve Dis.
, 8
(3
), pp. 312
–323
.10.
Ellis
, J. T.
, and Yoganathan
, A. P.
, 2000
, “A comparison of the hinge and near-hinge flow fields of the St Jude medical hemodynamic plus and regent bileaflet mechanical heart valves
,” J. Thorac. Cardiovasc. Surg.
, 119
(1
), pp. 83
–93
.11.
Travis
, B. R.
et al., 2002
, “An analysis of turbulent shear stresses in leakage flow through a bileaflet mechanical prostheses
,” J. Biomech. Eng.
, 124
(2
), pp. 155
–165
.12.
Meyer
, R. S.
et al., 2001
, “Laser Doppler velocimetry and flow visualization studies in the regurgitant leakage flow region of three mechanical mitral valves
,” Artif. Organs
, 25
(4
), pp. 292
–299
.13.
Lamson
, T. C.
et al., 1993
, “Relative blood damage in the three phases of a prosthetic heart valve flow cycle
,” ASAIO Trans.
, 39
(3
), pp. M626–M633
M626–M633
.14.
Travis
, B. R.
et al., 2001
, “Bileaflet aortic valve prosthesis pivot geometry influences platelet secretion and anionic phospholipid exposure
,” Ann. Biomed. Eng.
, 29
(8
), pp. 657
–664
.15.
Travis
, B. R.
et al., 2001
, “The sensitivity of indicators of thrombosis initiation to a bileaflet prosthesis leakage stimulus
,” J. Heart Valve Dis.
, 10
(2
), pp. 228
–238
.16.
Cockelet G., “The rheology of human blood” Fung Y.
17.
Pohl
, M.
et al., 1996
, “In vitro testing of artificial heart valves: comparison between Newtonian and non-Newtonian fluids
,” Artif. Organs
, 20
(1
), pp. 37
–46
.18.
Nygaard
, H.
et al., 1994
, “A new perivascular multi-element pulsed Doppler ultrasound system for in vivo studies of velocity fields and turbulent stresses in large vessels
,” Med. Biol. Eng. Comput.
, 32
(1
), pp. 55
–62
.19.
Nygaard
, H.
et al., 1994
, “Turbulent stresses downstream of three mechanical aortic valve prostheses in human beings
,” J. Thorac. Cardiovasc. Surg.
, 107
(2
), pp. 438
–446
.20.
Garbani
, J. L.
, Forster
, F. K.
, and Jorgensen
, J. E.
, 1982
, “Measurement of fluid turbulence based on pulsed ultrasound techniques
,” J. Fluid Mech.
, 118
, pp. 445
–470
.21.
Lieber
, B. B.
et al., 1989
, “On the discrimination between band-limited coherent and random apparent stresses in transitional pulsatile flow
,” J. Biomech. Eng.
, 111
(1
), pp. 42
–46
.22.
Benedict
, L. H.
, and Gould
, R. D.
, 1996
, “Towards better uncertainty estimates for turbulent statistics
,” Exp. Fluids
, 22
, pp. 129
–136
.23.
Tennekes, H., and Lumley, J. L. 1972, A first course in turbulence MIT Press, Cambridge, Mass.
24.
Grigioni
, M.
et al., 2000
, “On the monodimensional approach to the estimation of the highest reynolds shear stress in a turbulent flow
,” J. Biomech.
, 33
(6
), pp. 701
–708
.25.
Flachskampf
, F. A.
et al., 1991
, “Patterns of normal transvalvular regurgitation in mechanical valve prostheses
,” J. Am. Coll. Cardiol.
, 18
(6
), pp. 1493
–1498
.26.
Yoganathan
, A. P.
et al., “Oblique flow vectors from dispersing jets produce the velocity overestimation on angle corrected continuous wave Doppler studies: in vitro laser Doppler investigations
,” Circulation
, 76
(Supp IV), pp. 355
355
.27.
Blevins, R. D. 1984, Applied fluid dynamics handbook Van Nostrand Reinhold Co., New York, N.Y.
28.
Sahn
, D. J.
, 1988
, “Instrumentation and physical factors related to visualization of stenotic and regurgitant jets by Doppler color flow mapping
,” J. Am. Coll. Cardiol.
, 12
(5
), pp. 1354
–1365
.29.
Stewart
, S. F.
, Burte
, F.
, and Clark
, R. E.
, 1993
, “Quantification of asymmetric valvular regurgitant jets by color Doppler ultrasound in vitro
,” Echocardiography
, 10
(1
), pp. 23
–36
.Copyright © 2004
by ASME
You do not currently have access to this content.