Two different methods of attaching the current leads for segmental bioelectrical impedance analysis measurements were compared. The first method was leaving the current electrodes in place at the hand and foot for all measurements, while the second method involved moving the current electrodes to be adjacent to the voltage electrodes when measuring each segment. Each method was compared to a whole-body measurement performed at the same occasion as the segmental measurements. Slight deviation from the whole-body measurement was expected, as the electrode placements resulted in small gaps that were not measured between each of the arm, trunk, and leg segments. The second method, moving the current electrodes for each segment, was found to be the closer of the two methods for whole body measurements, with errors of 2% and 5.8% at R0 and R∞, respectively, while the first method had errors of 4.2% and 6% respectively.
Skip Nav Destination
2019 Design of Medical Devices Conference
April 15–18, 2019
Minneapolis, Minnesota, USA
ISBN:
978-0-7918-4103-7
PROCEEDINGS PAPER
A Comparison of Two Segmental Bioelectrical Impedance Analysis Methods With Whole-Body Analysis
Thomas Cannon,
Thomas Cannon
Georgia Southern University, Statesboro, GA
Search for other works by this author on:
JungHun Choi
JungHun Choi
Georgia Southern University, Statesboro, GA
Search for other works by this author on:
Thomas Cannon
Georgia Southern University, Statesboro, GA
JungHun Choi
Georgia Southern University, Statesboro, GA
Paper No:
DMD2019-3217, V001T10A005; 3 pages
Published Online:
July 19, 2019
Citation
Cannon, T, & Choi, J. "A Comparison of Two Segmental Bioelectrical Impedance Analysis Methods With Whole-Body Analysis." Proceedings of the 2019 Design of Medical Devices Conference. 2019 Design of Medical Devices Conference. Minneapolis, Minnesota, USA. April 15–18, 2019. V001T10A005. ASME. https://doi.org/10.1115/DMD2019-3217
Download citation file:
430
Views
0
Citations
Related Proceedings Papers
Related Articles
Statistical Investigation on Resistance Spot Welding Quality Using a Two-State, Sliding-Level Experiment
J. Manuf. Sci. Eng (August,2001)
Monitoring Resistance Spot Nugget Size by Electrode Displacement
J. Manuf. Sci. Eng (May,2004)
PEDOT:PSS/SWCNTs as a Transparent Electrode for Organic Photovoltaic Devices
J. Nanotechnol. Eng. Med (February,2012)
Related Chapters
Scope
Consensus on Operating Practices for the Sampling and Monitoring of Feedwater and Boiler Water Chemistry in Modern Industrial Boilers (CRTD-81)
Test Methods
Consensus on Operating Practices for the Sampling and Monitoring of Feedwater and Boiler Water Chemistry in Modern Industrial Boilers (CRTD-81)
Introduction
Ultrasonic Welding of Lithium-Ion Batteries